70 Automatic test call interval
You can program the interval between automatic test calls from the
Carephone 62 to the monitoring centre in days. In addition, there is a test
call after power up or leaving programming mode, as well as a test call per
randomization.
Select from 1 to 28 days.
0 = off
Default value is 0.
71 Action button (S button)
The Action button can be configured differently according to your needs:
0 = off
1 = service button / direct call (e.g. service call to a monitoring centre
or direct call to a relative)
3 = activate relay output (e.g. a door opener)
4 = send a manual test call (by pressing at least 2 seconds on the
action button)
Default value is 0.
72 Activate output
The Carephone 62 provides a potential-free relay output with a normally
open switch contact. The way the output reacts can be programmed.
0 = off
1 = speak / listen connection and repeated emergency call
3 = outgoing emergency call
4 = incoming call recognition
5 = wireless transmitter
6 = remote activation
7 = speak / listen connection
8 = pre-alarm
9 = pre-alarm and speak / listen connection
Default value is 0.