hit counter script

Acculab ATL-224 Operating Instructions Manual page 51

Acculab atilon electronic analytical and precision balances
Table of Contents

Advertisement

Specifications for Individual Models
Model
Weighing capacity
Readability
Tare range (subtractive)
Repeatability
(std. deviation)
Linearity
Response time (average)
Sensitivity drift within +10 to +30°C
Adaptation
to ambient conditions
External calibration weight
Operating temperature range
Net weight, approx.:
Weighing pan size
Weighing chamber height
Dimensions (W + D + H)
Model
Weighing capacity
Readability
Tare range (subtractive)
Repeatability
(std. deviation)
Linearity
Response time (average)
Sensitivity drift within +10 to +30°C
Adaptation
to ambient conditions
External calibration weight
(of at least accuracy class...)
Operating temperature range
Net weight, approx:
Weighing pan size
Dimensions (W + D + H)
ATL-224, ATL-224-I ATL-124, ATL-124-I ATL-84, ATL-84-I
g
220
g
0.0001
g
220
<± g
0.0001
<± g
0.0002
s
2.5
–6
2 · 10
<±/K
By selection of 1 of 4 optimized filter levels; display
update: 0.1 – 0.4 (depends on filter level selected)
g
200 (E2)
+10 ... +30 °C (273 ... 303 K, 50° ... 86 °F)
kg
4.4 | 4.7
mm
90 d
mm
230
mm
230 + 303 + 330
ATL-623,
ATL-623-I
g
620
g
0.001
g
620
<± g
0.001
<± g
0.002
s
1
–6
2 · 10
<±/K
By selection of 1 of 4 optimized filter levels; display
update: 0.05 – 0.4 (depends on filter level selected)
g
500 (E2)
+10 ... +30 °C (273 ... 303 K, 50° ... 86 °F)
kg
3.2 | 3.6
mm
115 d
mm
230 + 303 + 136
120
80
0.0001
0.0001
120
80
0.0001
0.0001
0.0002
0.0002
100 (E2)
50 (E2)
ATL-423,
ATL-153,
ATL-423-I
ATL-153-I
420
150
0.001
0.001
420
150
0.001
0.001
0.002
0.003
1
1.3
–6
2 · 10
5 · 10
200 (F1)
100 (F1)
3.2 | 3.6
2.6 | 3.0
–6
51

Advertisement

Table of Contents
loading