Page 2
Cert. No. Q5907 As a part of ABB, a world leader in process automation technology, we offer customers application expertise, service and support worldwide. We are committed to teamwork, high quality manufacturing, advanced EN 29001 (ISO 9001) technology and unrivalled service and support.
Startup ..........17 environment must taken into consideration. APPENDIX A – ENVIRONMENTAL PROTECTION ..........19 • The installation this equipment must be in accordance with APPENDIX B – MAGMASTER BLOCK relevant national and local standards. DIAGRAM ............20...
…3 ELECTRICAL INSTALLATION 3.2.4 Conduit Adapters and Cable Glands (North American – 0.5in) Alternative Alternative Face Seal Face Seal Face Seal Face Seal Ferrule Illustration Fittings vary Seal typical for slightly for (Present in O.Z.Gedney different makes O.Z. Gedney Fittings) Outer Nut Outer Nut Conduit Adapters...
Page 11
3 ELECTRICAL INSTALLATION… Caution. (Remote versions) • Remove any exposed black conductive layer from under coaxial screens. • Make connections only as shown. • Sleeve all bare wiring. • Twist RED and YELLOW cores lightly together. • Twist WHITE and GREY coaxial cables lightly together. •...
…3 ELECTRICAL INSTALLATION 3.3.2 Transmitters (All versions) Remove Protection Cover Caution. Unused cable entries must be blanked with Slide permanent down blanking plugs Pull out supplied with the slightly... transmitters. Slacken ...and captive slide off screws Fig. 3.12 Transmitter Connection Terminal access Caution.
Page 13
Fig. 3.14 Sensor Cable Connections at the Transmitter (North American Wiring Practice) 3.3.3 MagMaster-CalMaster Adapter When a MagMaster Transmitter is fitted with an adaptor board for use with a CalMaster Verification Unit, wiring from the sensor to this adaptor board is shown in the following diagram.
However, the supply connection in the former is made using a non-reversible plug (provided). 3.4.1 Frequency Outputs – Fig. 3.16 3.4.2 PLC Interface – Fig. 3.17 Counter/Totalisers MagMaster Transmitter d.c. Typically supply 1kΩ 1W Forward Flow d.c.
Page 15
• Operation of outputs is programmable – see Configuration Manual for details. • Frequency and Alarm outputs share a common return with contact input. • External isolators not normally required, as the pulse, alarm and contact circuits are electrically separated from all other Magmaster connections. Timer MagMaster Transmitter d.c.
Page 16
3.4.5 Current Output – Fig. 3.20 and 3.21 Information. • Output is fully programmable – see Programming Guide. • Output is electrically separated from all other MagMaster connections. • External isolators are not normally required and may significantly limit accuracy if used. MagMaster Transmitter HART link IC –...
Page 17
3 ELECTRICAL INSTALLATION… 3.4.6 Computer Connection – Fig. 3.22 and 3.23 Information. RS422/423 option is electrically isolated from all other MagMaster connections. MagMaster RS422 Connection APPLE Connector TERMINALS NAME (8 Pin MC) TX-SIG RX DATA - TX+SIG RX DATA +...
…3 ELECTRICAL INSTALLATION 3.4.7 Power Supply Connections – Fig. 3.24 and 3.25 Warning. • DISCONNECT THE SUPPLY FROM ANY CABLES BEING TERMINATED ON THE TRANSMITTER. • Electrical installation and earthing (grounding) must be in accordance with relevant national and local standards. •...
Quick Reference Programming Guide or the main MagMaster manual. 4.1 Startup Switch on the power supply to the flowmeter, and • For all versions supporting HART , see if a transmitter with display has been ordered, the the main MagMaster manual.
Page 20
…4 STARTUP AND OPERATION 12.328 Ltr/s >43567 MagMaster 32.8 Ltr/s >42315 Fig. 4.2 Location of Controls (Keypad Versions)
APPENDIX A – ENVIRONMENTAL PROTECTION Warning. • Potting materials are toxic – use suitable safety precautions. • Read the manufacturers instructions carefully before preparing the potting material. • The remote sensor terminal box connections must be potted immediately on completion to prevent the ingress of moisture.
Page 23
– Food & Beverage and Repair Centre. – Manufacturing – Metals and Minerals United Kingdom – Oil, Gas & Petrochemical ABB Limited – Pulp and Paper Tel: +44 (0)1453 826661 Fax: +44 (0)1453 829671 Drives and Motors • AC and DC Drives, AC and DC Machines,...