USER INSTRUCTION MANUAL OML Remote Model: OML Remote V1.0 Software: V1.0 Keep these instructions for future reference. Read the instructions and precautions before using this device. …a confident choice...
Operating Guide Key Features 18-21 Turning On & Off Battery Level Unlocking and Locking Options Menu 25-26 ® Joining an ODFS Pace XL Replacing or Rejoining a Stimulator Removing a Stimulator Wireless Footswitches (OML LINQ 30-31 Stimulator Battery, Changing Channel...
Page 3
Adjusting Intensity Testing Stimulation Active Mode Pause Mode Stimulation Mode Selection Charging the OML Remote Wireless Charging the OML Remote Attaching a Lanyard Screen Symbols, Software Version Caring for your OML Remote Maintenance, Servicing and Calibration Cleaning and Care Guidance...
Pace XL. ® Introduction The OML Remote can be used to control a maximum of 2 stimulators. It is powered using an internal rechargeable battery and can be carried using a lanyard (not supplied). The product is also capable of wireless charging which eliminates the need to plug in a charger by placing it on an appropriate charging mat (not supplied).
Introduction Intended Environment The OML Remote is intended to be used in the home, public and clinical environment. It is expected that users would store the OML Remote in their pocket, on a lanyard or in a handbag when out and about.
If using a lanyard, it is advised to use a lanyard with a saftey break release clip. Children The OML Remote is not a toy! Keep out of reach of children. Batteries All battery technologies carry a very small risk of fire or explosion if used incorrectly.
Page 8
Important Information (Warnings) Air Travel/Wireless Restricted Areas The OML Remote could cause radio-frequency (RF) interference to other critical equipment. Do not use the OML Remote in areas where radio frequency devices are prohibited (e.g. on aircraft). Flammable Cleaning Products Do not use flammable cleaning products to clean the device.
When charged in a warm environment above 36 °C the surface of the OML Remote and wireless charging mat may exceed 41 °C Do not handle the device until it has cooled.
Page 10
Water Ingress Do not get the OML Remote wet. Do not use (operate or charge) if the OML Remote is wet. Water ingress may stop the OML Remote from working. If submerged, turn off the device, dry out thoroughly and return to the manufacturer for assessment prior to re-use.
Page 11
Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the OML Remote including the stimulators joined to it and their accessories. Otherwise, degradation of the performance of this equipment...
Page 12
Do not touch the terminals of the charging socket. The ODFS Pace XL may not immediately unpause or pause ® upon pressing the pause button on the OML Remote. Wait for confirmation from the ODFS Pace XL or OML Remote prior to ®...
Read the instructions before use. Consult the instructions for use. Do not charge if the OML Remote is wet, the charger is wet, or your hands are wet. Warning, surfaces may be hot! This product should not be disposed of with other household waste.
Page 14
Do not get any part of the system wet. Expiry date (Year, Month, Day) of product (may relate to specific components). Do not use the OML Remote or accessories if the packaging is substantially damaged. Proceed with caution and check components for damage prior to use.
Page 15
Do not store outside the specified relative humidity limits. Do not store outside the specified atmospheric pressure limits. Batch number. Unique serial number of the OML Remote. Odstock Medical Limited device reference number. European authorised representative. Manufactured by Odstock Medical Limited: Odstock...
Definitions and Descriptions Lock/Unlock The buttons of the OML Remote will be locked (disabled) after 15 seconds of no use, to prevent accidental activation. Press the ‘standby’ button to unlock the buttons. If already unlocked press the ‘standby’ button if you want to lock the buttons before 15 seconds has elapsed.
See page 39 (Wireless Charging). Warranty The OML Remote is warrantied for a period of one year from the date of sale or fitting. Should any failure of the device occur during the warranty period, the device should be returned to Odstock Medical Limited for inspection.
Key Features Display Standby Channel Test Select Pause Down Mode Front View of the OML Remote...
Page 19
Displays information about the OML Remote and the stimulator(s) it controls. Standby Button Press and hold to power the OML Remote ON and OFF. Press to lock and unlock the use of the OML Remote. Pause Button Press to pause and activate stimulators remotely. If navigating a menu, press to select menu options.
Page 20
Key Features OML Logo Charger Alignment Guide Charging Socket Part Number Serial Number Lanyard Attachment Rear View of the OML Remote...
Page 21
Key Features OML Logo This marks the centre of the back of the OML Remote. Use it to align the OML Remote on a wireless charger. Lanyard Attachment Use to attach a lanyard. Charging Socket Plug in the charger to charge the battery.
“OML Remote” is displayed on the screen and fades away. Hardware Reset If the OML Remote is not responding, it may need to be reset. To do a hardware reset of the OML Remote, press and hold the standby button for 6 second until the screen turns off.
Battery Level The battery level of the OML Remote is displayed at turn on and when the OML Remote is unlocked (see page 24). Battery full (100%) Battery half-full (50%) Battery empty (<5%) When the battery level reaches battery empty the OML Remote must be charged.
To prevent accidental adjustment of the ODFS Pace XL, the OML ® Remote automatically locks its buttons if not used for more than 15 seconds. The OML Remote needs to be unlocked before it can be used to control the ODFS Pace XL. ®...
Options Menu The OML Remote includes a menu that allows the configuration of beeps, vibration and the wireless network. This can be accessed by pressing and holding the Mode button for approximately 2 seconds: Press the Mode button to cycle between menu options.
Page 26
Options Menu Beeps The OML Remote generates beeps when buttons are pressed to inform you that a button press has been registered. Enter the Options Menu and select the beeps option to turn beeps on or off; use the pause button to select this option.
Turn on OML Remote and enter Options Menu. Select the wireless option. If no stimulators have previously been joined it will start the joining process for stimulator 1. The OML Remote will wait for up to 2 minutes for the ODFS Pace XL to join.
Replacing or Rejoining a Stimulator Replacing one of the joined ODFS Pace XLs with a ® new one. Enter the Options Menu and select the wireless option. Select which stimulator you Select which stimulator you would like to replace. would like to replace. By pressing the pause button, stimulator previously...
To do this, follow the process below. Once an ODFS Pace XL is removed, it can no longer ® be controlled using the OML Remote. Select which stimulator you would like to remove. Press the pause button again or wait 2 minutes for the joining animation to time out.
ODFS Pace XL. ® To join a wireless footswitch, follow the instructions from the stimulator manual or the stimulator joining process sheet. The OML Remote will need to be unlocked to allow the stimulator to join to a wireless footswitch.
Page 31
‘secondary’ stimulator. If unpausing directly on the ODFS Pace XL (i.e. without using the ® OML Remote), the ‘primary’ stimulator will need to be unpaused first. The ‘secondary’ stimulator will not trigger wirelessly without the ‘primary’.
It is also used to show the stimulation strength percentage and battery level of the stimulator. Ensure that the OML Remote is unlocked. Only changes channel if set up with multiple channels of stimulation. Press the channel select button until the desired stimulation...
(pulse width) of stimulation. The Down button is used to decrease the intensity (pulse width) of stimulation. The intensity of stimulation is displayed on the OML Remote as a percentage and ranges from 0% to 100%. Ensure that the OML Remote is unlocked.
Testing Stimulation The Test button on the OML Remote is used to produce a short burst of stimulation to test the effect. This can be useful to check the response to stimulation before unpausing. Ensure that the OML Remote is unlocked.
Active Mode Press the Pause button of the OML Remote to start stimulation and observe the display to confirm that all joined stimulators have entered an active state. Ensure that the OML Remote is unlocked. Stimulator activated. Failed to activate...
Pause Mode The Pause button of the OML Remote is used to stop stimulation. Ensure that the OML Remote is unlocked. Stimulator activated. Stimulator paused.
(for example a walking mode and an exercise mode), press the Mode button to switch between these. Ensure that the OML Remote is unlocked. When multiple stimulators are joined, these screens indicate that only one stimulator has changed to exercise as the other stimulator does not have exercise mode enabled.
Charging the OML Remote Only use the supplied charger when charging the OML Remote with a wall charger. Charge the battery before use if the OML Remote has not been operated for an extended period of time. Charging Alignment Arrows...
Wireless Charging the OML Remote If using the wireless charging capabilities of the OML Remote, only use a Qi compatible wireless charger. Ensure the OML Remote is located in the centre of the wireless charger. Battery charging animation Battery fully charged Do not charge if device is wet.
Attaching a Lanyard You can attach a lanyard or wrist strap to the OML Remote by fixing it to the dedicated attachment on the back of the OML Remote as shown below: Only use a lanyard around the neck that has a safety...
No connection to stimulator. (Try again) Failed to activate. No footswitch! Waiting to join OML Remote Software Version To check the software version of the OML Remote, turn the OML Remote off. Hold the Channel Select button while turning the OML Remote on.
OML Remote or accessories they should be replaced. Turn off the OML Remote if it is not to be used for an extended period of time. This is achieved by pressing and holding the standby button as shown on the key features page.
If the OML Remote or accessories are very dirty or not easily cleaned, replace with new ones. OML Remote The OML Remote can be cleaned by wiping with a damp cloth or anti-bacterial wipes. Make sure that the OML Remote is turned off prior to cleaning.
Fault/Problem Solution/Action OML Remote does not switch on 1. Follow the instructions on how to turn on the OML Remote on page 2. Ensure the battery is charged (refer to page 38). 3. If none of the above works, contact the manufacturer (contact details are on the back of this manual).
Page 45
(wireless disabled). 3. Turn off stimulator and OML Remote and try again. 4. Re-join the stimulators to the OML Remote if the network has been lost on them. Refer to the sections on joining (page 27) and the stimulator...
Page 46
2. Follow the instructions on how to place the OML Remote on the charger on page 39. 3. Ensure that no metal objects (e.g. keys) are on the wireless charger next to the OML Remote.
Page 47
Troubleshooting Fault/Problem Solution/Action Stimulation mode symbol is missing 1. Switch the OML Remote OFF and or incomplete. ensure all stimulators are on and within range. Turn The OML Remote back on after a few seconds. 2. If the above does not fix the issue,...
Page 48
Troubleshooting Fault/Problem Solution/Action Stimulation intensity cannot be The mode on the OML Remote does adjusted and I get a screen similar not match the selected/available to one of these ones below: mode on the stimulator. This is the case if one of your stimulators...
Page 49
OML Remote network and re-join them (see page 28-29). Remember re-join any wireless footswitches in the system. Joined OML Remote to When a stimulator is joined to an OML stimulator but now wireless Remote a new network is created trigger does not work.
Technical Specification OML Remote Classification Class I medical electrical equipment, internally powered, portable and hand-held, continuous operation. No applied part(s). Expected service life 2 years of device Ingress protection Rating of IP22. Vertically falling drops have no rating harmful effects when the stimulator is tilted at any angle up to 15°...
Page 51
Technical Specification OML Remote Operational Temperature 5 °C to 40 °C (41 °F to 104 °F) Charging Wired: 5 °C to 40 °C (41 °F to 104 °F) Wireless: 5 °C to 30 °C (41 °F to 86 °F) Storage and Transport 1 months: -20 °C to 60 °C (-4 °F to 140 °F)
Page 52
Technical Specification Power Supply Unit Classification TR15RAM Series 15 W Medical Switching Adaptor. Meets 2 means of patient protection with continuous short circuit protection. Expected service life of 2 years device Ingress protection Do not get the device wet. rating Input 90~264 VAC Output...
Page 53
Technical Specification Power Supply Unit Operational Temperature 5 °C to 40 °C (41 °F to 104 °F) Storage and Transport -20 °C to 85 °C (-4 °F to 185 °F) Input 90~264 VAC Relative humidity 93% maximum, non-condensing Atmospheric 700 hPa to 1060 hPa pressure Device size 80.6 mm x 47.9 mm x 43.3 mm...
CFR Part 15 (USA) meeting the requirements for modular transmitter approval as detailed in the FCC public notice DA00.1407.transmitter. Quality of service The OML Remote is designed and tested to have a wireless message latency of less than 100ms depending on system configuration. Co-existence and...
Page 55
Please refer to troubleshooting section. Like all radio telemetry devices, the OML Remote and other OML wireless devices and accessories should not be used while flying. Turn off the OML Remote.
OML Remote Manufacturer’s declaration of electromagnetic emissions The OML Remote is intended for use in the electromagnetic environment specified below. The customer or user of the OML Remote should ensure that it is used in such an environment. Emission test...
Page 57
OML Remote Manufacturer’s declaration of electromagnetic immunity The OML Remote is intended for use in the electromagnetic environment specified below. The customer or user of the OML Remote should ensure that it is used in such an environment. Immunity test...
Page 58
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in which the OML Remote is used, exceeds the applicable RF compliance level above, the OML Remote should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orientating or relocating the OML Remote.
Page 59
OML Remote Manufacturer’s declaration of electromagnetic immunity continued The OML Remote is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the OML Remote can help prevent electromagnetic interference...