hit counter script

Warm-Up Phase - ABB GAA330-M Operating Instructions Manual

Emission monitoring system for marine applications
Table of Contents

Advertisement

52

Warm-up phase

Warm-up phase
End of the warm-up phase
Readiness, sample gas supply
Calibration
GAA330-M EMISSION MONITORING SYSTEM
The warm-up time is approx. 2 to 4 hours.
The warm-up phase can take longer if the analyzer system was not brought
to room temperature before the power supply was activated.
During the warm-up phase measurement values can be outside the ranges
specified in the data sheet.
The warm-up phase is over when the temperature and flow status messages
are gone and the measured value drift is acceptable. The latter depends on
the size of the measurement range.
At the end of the warm-up phase the analyzer system is ready for operation
and automatically activates the sample gas supply.
Calibration should only be started after the warm-up phase.
|
OI/GAA330-M-EN REV. C

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents