Copyright 2016 Ombg3000 - Rev1 - LANDTEC BIOGAS 3000 Operating Manual

Table of Contents

Advertisement

2) Press the right soft key to 'Exit' and return to the 'Calibration' menu.
3) If fitting a new cell or complete module, confirm the cell type and range,
and press the left soft key to 'Continue'.
Note: Alongside the new cell or complete module there will be a
calibration certificate containing the new slope and offset figures. Both of
these values need entering in to the appropriate fields in the above
screen.
© Copyright 2016
General Operating Instructions
Screen 30 - Fit New External Cell confirmation
Screen 31 - Input new Slope and Offset
Operating Manual
OMBG3000 – Rev1
115